Description:SeeweTek Permanent Magnet Material Testing System is a magnetic property measurement instrument designed to test standard samples with intrinsic coercivity below 1900 kA/m, in compliance with national standard GB/T 3217-2013 and international standard IEC 60404-5. An optional temperature rise unit enables measurements at room temperature up to 200°C for permanent magnet materials in various shapes including cylinders, blocks, rings, and tiles.
Detail
Parameter
Other

SeeweTek Permanent Magnet Material Testing System is a magnetic property measurement instrument designed to test standard samples with intrinsic coercivity below 1900 kA/m, in compliance with national standard GB/T 3217-2013 and international standard IEC 60404-5. An optional temperature rise unit enables measurements at room temperature up to 200°C for permanent magnet materials in various shapes including cylinders, blocks, rings, and tiles.


Product Features:

  • High accuracy and repeatability
  • Bilingual Chinese/English user interface for ease of use
  • Direct input of geometric dimensions based on sample shape
  • Automatic nonlinear correction of Hall probe
  • Automatic flux drift correction
  • Real-time display of H/B sampling waveforms during magnetization and testing
  • Includes magnetization curve, recoil permeability, and magnetic field calibration functions
  • Displays B&J-H hysteresis loops and magnetization curves, with automatic saving of test results
  • Allows comparison of J-H demagnetization curve clusters under different grades or temperature conditions
  • Test reports with print preview and automatic matching print output

Applicable Standards:

  • GB/T 3217-2013 Test methods for magnetic properties of permanent magnet (hard magnetic) materials
  • GB/T 24270-2009 Method for measurement of temperature coefficient of magnetic properties of permanent magnetic materials
  • IEC 60404-5 Magnetic materials - Part 5: Methods of measurement of magnetic properties of permanent magnet (hard magnetic) materials

Magnetizing Power Supply

ModelType 140Type 200Type 250Type 300
Output Power2 kW4 kW8 kW16 kW
Power SupplySingle-phase 220 V, 50 HzThree-phase 380 V, 50 Hz
Output Voltage±100 V±120 V±150 V±200 V
Output Current±20 A±40 A±60 A±80 A
Stability0.05%, overvoltage, overcurrent, short-circuit protection, operating environment: 0-40°C, 40-80% RH, non-condensing


Fluxmeter Technical Parameters (for measurement of H and J in rare earth permanent magnets)

Analog output per range0 to ±5 V
Input impedance10 kΩ
Error0.5%
Drift1 μWb/60 s
Range2.5, 5, 10, 20 mWb (auto-ranging; range automatically calculated based on material type and geometry)
Drift adjustmentManual


Gaussmeter Technical Parameters (for measurement of H in general permanent magnets)

Analog output per range0 to ±5 V
Hall element temperature coefficient0.06%/°C
Software calibration accuracyBetter than 0.3%
Imported Hall chip probe (linear correction range)0 to 2.5 T
Range0.5/1.0/2.0/4.0 T
Error0.5%
Resolution1 mT
Linearity±0.2% (computer corrected)
Display4 1/2 digit LED
*The Hall probe maintains horizontal level during vertical and horizontal movement.


Heating Power Supply (Optional)

Control range50 to 220°C
Temperature sensors2
Resolution1°C
Heating rods100 W, 4 pcs
Temperature uncertainty±1°C
Heating power input220 V/5 A
Heating power output400 VA
Other non-standard accessories can be customized.


Data Acquisition Card

Magnetization time30 to 120 seconds (user-selectable based on proficiency)
Resolution and linearity16 Bit ±1/2 LSB; sampling frequency: 110 kHz
AD samples H and B from gaussmeter and fluxmeter analog outputs respectively.
Conversion time between B and H: < 25 μs, synchronization effect negligible; software determines Ba and Ha.


Electromagnet Parameters

ModelDCT-140DCT-200DCT-250
Pole diameter140 mm200 mm250 mm
Pole face (reduced)100 mm60 mm80 mm
Adjustable gap60 mm80 mm80 mm
Resistance≤ 5 Ω≤ 3 Ω≤ 2.5 Ω
Weight≈ 200 kg≈ 800 kg≈ 1400 kg
*Note:
Choose different electromagnets based on sample size or coercivity;
SeewTek offers electromagnet customization and magnetic field simulation services.


Measurement Accuracy:

ParameterBrHcBHcJ(BH)maxHkHmax
Uncertainty (%)1.01.01.52.01.01.0
Repeatability (%)±0.3±0.3±0.3±0.5±0.5±0.5
Note: Measurement accuracy traceable to national accredited laboratory samples. Standard AlNiCo sample tested at constant temperature 23°C.


Software Test Interface